3
– Operation
Thermo Scientific
Product Manual for CarboPac PA200 Columns
Page 16 of 37
065734-01
For Research Use Only. Not for use in diagnostic procedures.
3.
Operation
3.1
Dionex CarboPac PA200 column Operational Parameters
pH
0-14
Temperature limit:
4-60 °C
Pressure limit:
4500 psi
Organic Solvent Limit: 100% compatible
Typical Eluents:
potassium hydroxide, potassium methanesulfonate
Standard Flow Rate:
1 mm: 0.063 mL/min
0.4 mm: 0.010 mL/min
3.1.1
The Best Operational Guidelines
3.1.1.1
ALWAYS…
Use dedicated glass- and plastic- ware for sample handling.
Use high purity water (>18.2 MΩ-cm resistivity).
Keep your water eluent bottle blanketed with ~ 5 psi helium or nitrogen. Use new filtered
water if left un-blanketed for more than 30 minutes.
3.1.1.2
NEVER…
Proceed to a next installation step if the previous step has failed.
Start an installation with any of the check list items below missing.
Use ‘communal’ filtration units or filters made of unknown or unsuitable (e.g., cellulose
derivatives) materials.
Use MeOH or other organic solvents as rinse fluid in the autosampler. Use only water,
replaced daily, or use sampler wash bottles blanketed with ~3 psi Nitrogen or helium.
Run above 60 °C or 5000 psi.
3.1.2
Initial Check List
The following items MUST be available in your lab. The absence of any of these may compromise
your analysis.
Laboratory water unit delivering >18.2 megohm-cm water at the installation site.
Vacuum system for eluent vacuum filtration
Inert gas cylinder (helium or nitrogen) with a regulator valve (for example, a 0-200 psi
gauge on the low pressure side) and the appropriate size adaptors plus tubing
Plastic eluent bottles with gas-tight cap-fittings.