Thermo Scientific
10-1
10
Troubleshooting
Table 3. Furnace Troubleshooting.
Troubleshooting procedures involve working with high voltages
which can cause injury or death. Troubleshooting should only be
performed by trained personnel. This section is a guide to
troubleshooting controller and furnace problem.
Problem
Solution
Absence of current and presence of voltage on
element input terminal.
Heating element is cracked or open.
Replace broken element.
Heating element curls towards wall
1. Shut off immediately.
2. After the furnace cools, relocate the
element farther from wall.
Controller actual furnace temperature display
decreases when furnace is heating
Check thermocouple polarity. Red is always
negative (Refer to Section Troubleshoot-
ing).