Thermo Scientific
4-1
4
Installation
Do not exceed the electrical and temperature ratings printed on the dataplate of
the furnace.
Location
Install the furnace in a level area free from vibration with a minimum of two feet of
space above the unit and six inches of space, for air flow, around the unit. Allow 13
inches in front and 15 inches to the side of the furnace for the door to swing.
Power Wiring
For detailed wiring information, refer to the wiring diagrams at the end of this
manual.
The 1700 °C Box Furnaces are designed for operation on a power source of either
240 or 208 volts, single phase, at 50 or 60 Hertz. If operating on 208 volts, change
the wiring on the transformer from 240 volts to 208 volts for best performance.
Power wires, ground wires, and main disconnect or circuit breakers are not
provided with the 1700 °C Box Furnaces. Two holes located at the bottom rear of
the furnace (refer to
), provide access for the power
connection wires.
Improper operation of the furnace could result in dangerous
conditions. To preclude hazard and minimize risk, follow all
instructions and operate within design limits noted on the
dataplate.
Summary of Contents for BF51634BC-1
Page 1: ......
Page 6: ...List of Figures ii Thermo Scientific...
Page 20: ...Safety Notes Standards and Directives 1 14 Thermo Scientific...
Page 48: ...Communication Option Auto Tuning the Controller 7 10 Thermo Scientific...
Page 64: ...Troubleshooting 10 4 Thermo Scientific...
Page 65: ......
Page 66: ......
Page 69: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable...
Page 70: ...Error Codes 12 2 Thermo Scientific...
Page 72: ...WEEE Compliance 13 2 Thermo Scientific...
Page 73: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable...
Page 74: ...Spare Parts and Accessories 14 2 Thermo Scientific...