Operation - UP550 Controller
Local Mode Operation
6-4
Thermo Scientific
Local Mode Operation
To operate the UP550 controller in the LOCAL mode, proceed as follows:
1. Prom the RESET mode (RST light is lit on front panel), press the MODE key
once.
2. In the lower display, you will see "LOC: ON" with the word "changing" flashing.
Press the SET/ENT key. The LOC light is now lit indicating you are in the
LOCAL mode.
3. Use the up and down arrow keys to enter a setpoint. Press the SET/ENT key to
register the setpoint.
4. While in the local mode, it is recommended to enter an alarm value. The alarm
value should be approximately 10 °C above the setpoint. To enter an alarm
value, hold the SET/ENT key down for 3 seconds.
5. PROG will appear in the lower display. Press the "up arrow" key once. “AL” will
appear. Press the SET/ENT key once then enter the alarm value using the
up/down arrow keys. Press the SET/ENT key to register the alarm value.
The shut off the control's output, place the unit in RESET mode by holding down
the RESET key for 2 seconds.
Summary of Contents for BF51634BC-1
Page 1: ......
Page 6: ...List of Figures ii Thermo Scientific...
Page 20: ...Safety Notes Standards and Directives 1 14 Thermo Scientific...
Page 48: ...Communication Option Auto Tuning the Controller 7 10 Thermo Scientific...
Page 64: ...Troubleshooting 10 4 Thermo Scientific...
Page 65: ......
Page 66: ......
Page 69: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable...
Page 70: ...Error Codes 12 2 Thermo Scientific...
Page 72: ...WEEE Compliance 13 2 Thermo Scientific...
Page 73: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable...
Page 74: ...Spare Parts and Accessories 14 2 Thermo Scientific...