0.2 Micron Final Filter Replacement
Replace the 0.2 micron final filter whenever any of the fol-
lowing conditions occur: the product water flow rate is
reduced or you experience bacteria break through, when
cartridges are replaced, or when system is cleaned. The
0.2 micron final filter is shipped assembled with a bell. To
replace the 0.2 micron final filter assembly:
1. Remove the old 0.2 micron final filter assembly
by turning it to the left to unscrew it from the
draw-off valve.
2. Remove the new 0.2 micron final filter assem-
bly from its bag and insert it into the draw-off
valve. Gently tighten, turning the filter to the
right.
3. Open the draw-off valve and flush at least 3
liters (1/2 reservoir volume) of water through
the 0.2 micron final filter.
Ventgard Cap Replacement
The purifying media and filter in the Ventgard filter ele-
ment have a limited capacity. Therefore, the Ventgard cap
should be replaced every 120 days. The Ventgard cap is
shipped as a complete unit; replacement involves simply
removing the new Ventgard cap from its plastic storage
bag and placing it on the reservoir. A Ventgard cap can be
stored in a cool, dry place for two years, provided its plas-
tic storage bag has not been opened.
Fuse Replacement
1. Turn off the EASYpure R
O
D
I
and disconnect it
from the power supply. Remove the power cord.
2. Pull out the fuse drawer located in the power
entry module.
3. Remove old fuses and replace with fuses of
the same type and rating. (See
Replacement
Parts
section.)
4. Replace fuse drawer.
39
M
AINTENANCE AND
S
ERVICING
Note
It is suggested th
at PTFE
tape be
applied to the threads of the 0.2 micron
final filter to ensure a tight seal.
Warning
Replace fuses with those of the same
type and rating.