Cleaning the Resistivity Cell
(See Fig. 13)
1. Turn off the EASYpure R
O
D
I
and disconnect it
from the power supply. Remove the power
cord.
2. Depressurize the system by opening the unit
dispenser draw-off valve, allowing water to
drain until no more flows from the valve.
3. Remove the Ventgard cap.
4. Remove the screws securing the EASYpure
R
O
D
I
top cover.
M
AINTENANCE AND
S
ERVICING
Warning
Depressurize system prior to opening
cartridge access door.
Note
Ensure you have an o-ring (part #
GSX29) available prior to cleaning
resistivity cell.
Figure 13: Disconnecting Resistivity Cell