4-43 AutoXP Users Guide
HART
5. AutoXP will modify the output to provide 4mA.
Figure 4-79. Fld dev output equal to reference meter
If the output goes to 4mA, proceed to step 6 below.
If the output is not 4mA, go back to step 4.
6. 20mA is forced onto the current loop.
Figure 4-80. Setting fld dev output to 20mA