5. Safety precautions
For your own safety, please observe the above safety precautions!
Pacific AFT system is a modularly constructed, pure water system that serves exclusively
for the purification of tap water.
Do not put the system into operation until you have taken notice of all of the appropriate
information that is given in these Operating Instructions.
Lifting and carrying the pure water system, e.g. to the installation location, should be
carried out by two people. To lift it, each takes hold of it under the base plate at two
corners.
Note that the manufacturer is freed of all liability for damages that result from improper
operation of the system, or from use of it for other than the intended purpose.
The CE-Mark becomes invalidated should constructional changes be made to the system
or products of other manufacturers be installed in it.
Protect the system from frost. The temperature in the area in which the system is installed
must be at least +2° C and must not 40° C.
Observe all regulations and requirements, including current accident regulations, that are
applicable and appropriate at the installation area, including those for the statics of the
flooring (see weight under „Technical specifications).
The raw water pressure must be at least 2 bar and at most 6 bar, should it be higher, then
an additional pressure reducer must be installed.
DIN EN 1717 requires that water purification systems be equipped with a safety device
that protects against contamination of the drinking water piping.
An earthed 100-250V / 50/60Hz socket must be available.
The installation area must have a drain at floor level with at least DN 50 pipe, otherwise
the manufacturer will not accept any liability for water damage.
Gravity fall to the waste drain must be ensured.