Pure water system
47
17. Emergency supply
Normal use
Emergency
supply
Normal use
Emergency supply
Right side of
Pacific-AFT
Left side of
the tank
Should the system or the tank show a fault, then the supply of purified water can be maintained
by appropriate utilization of the filter cartridge that is installed in the system.
09 Switch off the system and close off the supply of feedwater.
09 Turn the emergency supply tap to the „Emergency supply“ position.
09 Change the plugging of the quick connect coupling on the tank from „Normal use“ to
„Emergency
supply“.
09 Switch the system on and re-open the supply of feedwater.
5.
Use pressure reducer V9 on the tank to adjust the water pressure to the user to the
permissible value (2 bar), if appropriate. Turning anti-clockwise reduces the pressure and
turning clockwise increases it. Pressure gauge PI 200 shows the user water pressure.
Important:
Do not forget to re-adjust the pressure on returning to normal use.
The system can only be used for a limited time in emergency mode, i.e. until the capacity of the
filter cartridge is exhausted.
Two filter cartridges should always be kept available as reserve.
Please inform Service immediately when you switch your Pacific-AFT to emergency supply.