T
Thermo Fisher Barnstead
Pacific AFT
Pure water system
37
14.3. Replacing the RO membrane
Pacific 3, 7, 12 AFT:
1 RO-Membrane
Pacific 20 + 40 AFT:
2 RO-Membranes
1
2
3
4
Ansicht von hinten - ohne Rückwand
Back view, with back panel removed
Disconnect the line plug
Undo all hose connections to the pressure tube (1) of the reverse osmosis module.
Remove the pressure tube from the holding clamps.
Open cap nut (3) on the pressure tube and draw out the reverse osmosis membrane (2).
Insert the new reverse osmosis membrane in the pressure tube, with the two O-rings on
the permeate tube (4) foremost as indicated by the arrow.
Install the reverse osmosis module in the reverse succession.
Incorrect insertion of the reverse osmosis membrane would result in
immediate damage to it.