T
Thermo Fisher Barnstead
Pacific AFT
Pure water system
34
13.2.11 Adjusting the circuit hysteresis of the float switch:
An eleventh press on the menu-key:
Basic setting:
Off: 100 %
On:
70
%
Setting range:
Off:
25 – 100 %
On:
0 – 70 %
With a setting over 100 % for the upper switching point, the display of the tank level is switched
off. The setting here is so according to whether an analogue or a digital float switch is installed.
The display shows:
OEM-Menu
Off: 100 %
On: 70%
13.2.12 Programme choice, TII/TII UV:
A twelfth press on the menu-key:
In this menu, the equipping grade of the system can be set, to differentiate between UPW
and PW.
Basic setting: TII UV
The display shows:
OEM-Menu
Version
TII UV
13.2.13 Entering the type and serial number of the system:
In this menu, the type and serial number of the system can be entered, both of which are
then given as headline on every print-out. The following types of systems can be given:
Pacific-RO, Pacific-TII, Pacific-TII UV, Pacific-AFT.
The display shows:
OEM-Menu
Type Pacific AFT
S.Nr.: 9999/04