Thermo Fisher Barnstead
Pacific AFT
Pure water system
19
11. Bringing the system into service
1.
Use the R ¾“ hose supplied to connect the feedwater inlet
connector of the system (labelled „
raw water
“) to the prefilter
outlet. Use a further hose (R ¾“ hose, accessory for the pre-
treatment system) to connect the prefilter inlet to the closed
water tap.
2.
Use the second R ¾“ hose supplied to connect the
„Concentrate“
outlet of the system to the on-site drain.
Important!
The concentrate must be able to flow to drain by
free gravity fall. The drain to the stewer must be max. Are 1 m
above the rinsing water connector of the unit.
09 Use the 6 mm diameter hose supplied to connect the “
Recirculation
” connector of the
system to the “
Recirculation
” connectot of the tank.
09
Fit the quick connect coupling in the end of the 8 mm hose supplied and
fasten it on with the hose clip. Fit the coupling onto the lower tank connector for normal
use and listen to ensure that it clicks into position. Connect the other end of this hose to
the “Purfied water” connector of the system.