Chapter 7: Maintenance
iEMS
Incubator/Shaker User Manual
Rev. 3.0, Cat. no. 1506200A
36
7.2
How to replace fuses
The fuses are located on the left side of the instrument (Fig. 4.1, item 8).
1. Switch the instrument OFF. Unplug the instrument and disconnect the
power cable (cord) from the power input socket.
2. Use a small flat screwdriver to pry the voltage selector/fuse block assembly
away from the power entry port housing and remove (Fig. 7.1).
Fig. 7.1 Opening the voltage selector/fuse block assembly
3. Replace the blown fuses with the same type (3.15 A). See Fig. 7.2.
Fig. 7.2 Fuse block assembly
4. Place the fuse holder back into the slot and push it in until you hear a click.
5. Connect the power cable to the instrument and switch it ON.
2
1
1. Fuses
2. Voltage selector