Chapter 6: Operation
iEMS
Incubator/Shaker User Manual
Rev. 3.0, Cat. no. 1506200A
19
6 Operation
6.1 Principle
of
operation
When you power on the instrument, it will be in the READY state. From the
READY state you can choose to modify the setup parameters (Section 5.3.3
Setup) or the protocol parameters (Section 6.2 Programming protocols). If you
have already programmed protocols, while in the READY state, you can start to
preheat the required slots (Section 6.5 Preheating and heating) before starting
the operation (Section 6.4 Normal operation). It is also possible to start shaking
the slots without heating as a separate function (Section 6.6 Shaking). For an
overview of the operations, see Fig. 6.1 below.
Fig. 6.1 Principle of operation
Power ON
READY
shake
start
Shaking only
SHAKING
ON
READY
Change protocol
parameters
Slot no.
1 – 9
or
0
for all slots
↵
(Enter)
Set protocol
parameters,
see 6.2
READY
Start
protocol
Slot requires
preheating
start
READY
Protocol
starts
temp
start
HEAT ON
READY
Slot no.
1 – 9
or
0
for all slots
Slot no.
1 – 9
or
0
for all slots
Slot no.
1 – 9
or
0
for all slots
Summary of Contents for iEMS
Page 1: ...iEMS Incubator Shaker Rev 3 0 Cat no 1506200A iEMS Incubator Shaker User Manual ...
Page 59: ...iEMS Incubator Shaker User Manual Rev 3 0 Cat no 1506200A 59 ...
Page 60: ...iEMS Incubator Shaker User Manual Rev 3 0 Cat no 1506200A 60 ...
Page 61: ...iEMS Incubator Shaker User Manual Rev 3 0 Cat no 1506200A 61 ...