Chapter 5: Installation
iEMS
Incubator/Shaker User Manual
Rev. 3.0, Cat. no. 1506200A
15
Retain the original packaging and packing material for future transportation. The
packaging is designed to assure safe transport and minimize transit damage.
Use of alternative packaging materials may invalidate the warranty. Also retain
all instrument-related documentation provided by the manufacturer for future
use.
5.2.2
Checking delivery for completeness
Check the enclosed packing list against order. In case of any deviations, contact
your local Thermo representative.
5.2.3
Checking for damage during transport
Visually inspect the transport package, the instrument and the accessories for
any possible transport damage.
If the carton has been damaged in transit, it is particularly important that you
retain it for inspection by the carrier in case there has also been damage to the
instrument.
Neither the manufacturer nor its agents can be held responsible for any damage
incurred in transit, but the manufacturer will make every effort to help obtain
restitution from the carrier. Upon receipt of the carrier's inspection report,
arrangements will be made for repair or replacement.
Visually check all interconnections in the basic instrument. Check that there are
no loose parts inside the instrument.
If any parts are damaged, contact your local Thermo representative.
5.2.4 Climatic
requirements
When you set up your iEMS Incubator/Shaker, avoid sites of operation with
excess dust, vibrations, strong magnetic fields, direct sunlight, draft, excessive
moisture or large temperature fluctuations.
•
Make sure the working area is flat, dry, clean and vibration-proof and leave
additional room for accessories, cables, reagent bottles, etc.
•
Leave sufficient space (at least 10 cm) on both sides and at the back of the
unit to allow adequate air circulation.
•
Make sure the ambient air is clean and free of corrosive vapors, smoke and
dust.
•
Make sure the ambient temperature range is b10°C (50°F) and
+40°C (104°F).
•
Make sure relative humidity is between 10% and 90% (non-condensing).
The iEMS Incubator/Shaker does not produce operating noise at a level that
would be harmful. No sound level measurements are required after installation.
Warning:
Do not operate the instrument in an environment where potentially
damaging liquids or gases are present.
Summary of Contents for iEMS
Page 1: ...iEMS Incubator Shaker Rev 3 0 Cat no 1506200A iEMS Incubator Shaker User Manual ...
Page 59: ...iEMS Incubator Shaker User Manual Rev 3 0 Cat no 1506200A 59 ...
Page 60: ...iEMS Incubator Shaker User Manual Rev 3 0 Cat no 1506200A 60 ...
Page 61: ...iEMS Incubator Shaker User Manual Rev 3 0 Cat no 1506200A 61 ...