When setting up, installing and operating the unit, be sure to comply with all applicable regulations
in your country. (FRG: ZH 1/119, ZH 1/342, ZH 1/343, ZH 1/598)
Transport
Handle the unit with care during transport. Do not lift up the unit by its door.
Refer to TECHNICAL SPECIFICATIONS for weights and dimensions.
Place of installation
The place of installation must be dry and draught-free.
The unit should not be placed or operated in recesses which cannot be ventilated.
To avoid or at least minimize deviations from the technical specifications, the tempera-
ture at the place of installation must be in the range +18°C to +30°C.
Set up the unit on a solid, non-flammable surface (laboratory tables, subframes).
Make sure that the unit is in a stable and perfectly upright position.
Do not cover or obstruct the fresh air or exhaust vents in the unit housing
(please observe the minimum clearances - see page 11).
CO
2
are admitted into the gassed incubator. CO
2
is a health risk hazard. The place of installa-
tion must therefore be well-ventilated.
Gases discharged from the pressure relief valve must be safely expelled outside by means of the
ventilation systems.
It may be necessary to take special measures in order to ensure proper ventilation when several
units are installed in the same room.
For information on gas flow during operation, refer to the APPENDIX.
Stacking
The unit is not suitable for stacking when operated in the manner intended.