![Thermo Electron 3-Star Conductivity User Manual Download Page 95](http://html.mh-extra.com/html/thermo-electron/3-star-conductivity/3-star-conductivity_user-manual_1104301095.webp)
IX-23
Orion Star Series Meter
Thermo Electron Corporation
Key Feature Overview
Error Codes & Debugging
If an error code is displayed on the screen, the user must press the
MEAS key to clear.
Error/Event codes will be in the format “E###”. Some of these are errors, some are
warnings, and some are purely informational. The first digits identifies the type of
error or event as identified in the table below:
System Events/Errors
Description
Error Code
Description
Error Code
System Event or Error
E0##
Calibration Event or Error
E7##
pH Event or Error
E1##
Setup Event or Error
E8##
ISE Event or Error
E2##
SW Event or Error
E9##
ORP Event or Error
E3##
HW Event or Error
EA##
Conductivity Event or Error
E4##
Memory Event or Error
EB##
DO Event or Error
E5##
ISE Event or Error
EC##
Measurement Event or Error
E6##
System Events/Errors
Error
Code
Description
Press
MEAS
to Clear
Cause
Trouble Shooting
E001
No Model Number
Yes
Memory Corrupted
Call Customer Service
E002
No HW Calibration
Yes
Memory Corrupted
Call Customer Service
E003
Bad HW Calibration Yes
Memory Corrupted
Call Customer Service
E004
RTC Failure
Yes
HW Transeint or
Failure
Reset Meter, if error reoccurs
contact Customer Service
E005
ADC Failure
Yes
HW Transeint
or Failure
Reset Meter, if error reoccurs
contact Customer Service
Summary of Contents for 3-Star Conductivity
Page 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Page 20: ...Thermo Electron Corporation II 8 Orion Star Series Meter ...
Page 38: ...Thermo Electron Corporation IV 6 Orion Star Series Meter ...
Page 72: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter ...
Page 102: ...Thermo Electron Corporation X 6 Orion Star Series Meter ...
Page 108: ...Thermo Electron Corporation XI 6 Orion Star Series Meter ...