IV-3
Orion Star Series Meter
Thermo Electron Corporation
pH
pH Calibration for 1-5 Points
1. Prepare the electrode for use according to the instructions received with the
electrode.
2. Select the buffer set of NIST or DIN being used for the proper buffer recognition
to occur.
Note:
the calibration points should bracket the expected pH range of the
samples to be measured. These should be not less that 1 pH unit apart and not
more than 4 pH units apart.
▲
GLP indicates that at least a two-point calibration should be performed before
the first samples of the day are analyzed and a single point calibration should be
performed every two hours.
Fresh buffers should be used each time you calibrate.
Care should be taken when rinsing the electrode with deionized water and
GENTLY BLOTTING all remaining water drops with a lint-free tissue prior to
moving the electrode to the next beaker. DO NOT wipe a glass or epoxy
electrode body with a tissue as static build up will result.
3. Place the electrode and ATC probe into the first buffer. The neutral buffer
(pH 7.01 or 6.86) is usually recommended to be used first followed from lowest
to highest pH.
Summary of Contents for 3-Star Conductivity
Page 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Page 20: ...Thermo Electron Corporation II 8 Orion Star Series Meter ...
Page 38: ...Thermo Electron Corporation IV 6 Orion Star Series Meter ...
Page 72: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter ...
Page 102: ...Thermo Electron Corporation X 6 Orion Star Series Meter ...
Page 108: ...Thermo Electron Corporation XI 6 Orion Star Series Meter ...