IX-9
Orion Star Series Meter
Thermo Electron Corporation
Key Feature Overview
General Instrument Setup Menu
The following table describes the Measurement Setup Options:
General Instrument Setup
Description
Display
Range of Values
Default
Value
Method
Specific
Manual Temperature
Compensation Setting
gEn
dEgC
25.0
-5.0 - 105
25.0
Yes
Stirrer Speed Setting
gEn
StIr
4
OFF, 1, 2, 3, 4, 5, 6, 7
4
Yes
Auto-Shut Off Selection
dUE
ISE
0
gEn
AUtO
On
off, on
Yes
•
Manual Temperature Setting:
The is the temperature that will be used
for temperature compensation when no temperature sensors are attached to
the instrument.
•
Stirrer Speed Setting:
This screen is only present on the Benchtop meters.
It is used to set the stirrer speed. The speeds are 1 through 7 with speed 1
being the slowest qand speed 7 being the fastest. We generally recommend a
speed of 3 or 4 for most applications. When a stirrer speed is selected:
• In continuous or timed reading mode the stirrer is toggled on and off with the
stirrer button.
• In AUTO-READ mode pressing the MEASURE key will start the stirrer and the
stirrer automatically turn off once the measurement is stable.
• When the AUTO-STIR BOD probe and AUTO-READ mode is selected, pressing
the button on the AUTO-STIR BOD probe will act just like pressing the
MEASURE key. It will turn on the stirrer, then when the reading stabilizes it will
automatically turn off the stirrer, freeze the display and print & log the results.
Summary of Contents for 3-Star Conductivity
Page 1: ...Orion Star Series Meter User s Guide Analyze Detect Measure Control ...
Page 20: ...Thermo Electron Corporation II 8 Orion Star Series Meter ...
Page 38: ...Thermo Electron Corporation IV 6 Orion Star Series Meter ...
Page 72: ...Thermo Electron Corporation VIII 14 Orion Star Series Meter ...
Page 102: ...Thermo Electron Corporation X 6 Orion Star Series Meter ...
Page 108: ...Thermo Electron Corporation XI 6 Orion Star Series Meter ...