Hardware Manual COMBILOG 1020
135
Diagnostic
Description:
By this command a diagnostic telegram will be sent to the Data-
logger. If the telegram has been received in correct form the mod-
ule will send this telegram back unchanged (echo telegram).
request telegram
ADR FNR SUBFCT
DATA
CRC
08 00 00 A5 37 MSB
LSB
response telegram
ADR FNR SUBFCT
DATA
CRC
08 00 00 A5 37 MSB
LSB
ADR ............ module address (hex 00..7F)
FNR ............ function number (hex 08)
SUBFCT ..... subfunction number (hex 0000)
DATA .......... dataword (hex A537)
CRC ............ check sum
CRC polynom: u
15
+ u
13
+ 1
CRC start value: hex FFFF