ADS1298RECG-FE Input Signals
39
SBAU181B – March 2011 – Revised Janurary 2016
Copyright © 2011–2016, Texas Instruments Incorporated
ADS1298R
4.8.1
RLD Common Mode Voltage
The RLD common mode voltage can be set to (AVDD+AVSS)/2 or to an externally provided source. If the
application requires the common mode to be set to any voltage other than mid-supply, this can be
accomplished by setting the appropriate bit in the
. On the ADS1298ECG-FE, the
external RLDREF voltage is set using resistor R1 and adjustable resistor R2 (R1 and R2 are not installed
by default).
During power-up, the firmware configures the device for internal RLDREF operation. To configure the RLD
circuitry manually, use the following steps and the controls found on the
ADC Register
tab.
1. Verify that the
Channel Input
is set to the
Normal Electrode
mode for all channels (
Channel
Registers
tab).
2. In
CONFIG3
control register (
Channel Registers
tab):
•
Enable the RLD Buffer (
RLD Buffer Power
=
Enable
)
•
Set the internal RLD reference (
RLDREF Signal Source
)
3. Select the electrodes for the RLD loop from the
Right Leg Drive Derivation Control Registers
controls
(
LOFF and RLD
tab)
Once these steps are completed, measure and verify that the voltage on either side of R38 is close to
mid-supply. This measurement confirms whether the RLD loop is functional.
The on-chip RLD signal can be fed back into the ADS1298R by shorting JP1. This RLD signal can then be
sent to the ADC (to measure for debug purposes) or to other electrodes for driving (to change the
reference drive in case the RL electrode falls off). Refer to the
for
additional details.
4.8.2
Driving the RLD Cable Shield
Apart from the RLD signal, the ADS1298RECG-FE also offers three options to drive the cable shield:
•
In-phase RLD signal
•
Out-of-phase RLD signal
•
Board AGND
summarizes the configuration of JP15 and JP17 for each of the options.
Table 4. RLD Jumper Options
ECG Cable
ELEC_SHD signal
JP15
JP17
AGND
1-2
Don't Care
RLD (0: In phase)
2-3
2-3
RLD (180: Out of phase)
2-3
1-2
4.9
PACE Detection
The ADS1298R supports data rates up to 32kSPS for software PACE detection, which typically requires a
data rate of at least 8kSPS.
NOTE:
The ADS1298RECG-FE does
not
include software PACE detection algorithms.
The ADS1298R provides the user the flexibility of doing hardware PACE detection with external circuitry.
PACE detection can be done simultaneously on two channels: one from the odd channels and one from
the even channels. Refer to the
for additional details.
To turn on the PACE buffer and select the channels, set the
PACE
Register from the
GPIO and OTHER
Registers
tab. The PGA outputs of the selected channels are available at connector J5, pins 1 and 2.