
Temperature ( C) =
°
Temperature Reading ( V)
145,300 V
m
-
m
490 V/ C
m °
+ 25 C
°
2x
1x
1x
8x
AVDD
AVSS
ADS1298RECG-FE Input Signals
36
SBAU181B – March 2011 – Revised Janurary 2016
Copyright © 2011–2016, Texas Instruments Incorporated
ADS1298R
4.3
Temperature Sensor
The internal temperature sensor on the ADS1298R is shown in
. When the internal MUX is
routed to the temperature sensor input, the ADC internal temperature is calculated from the ADC output
voltage using
.
Figure 32. Internal Temperature Sensor
(1)
The ADC can be configured to give a temperature reading by selecting the
Temperature Sensor
option on
the Channel Control Registers GUI (see
) and verified using the
Scope
tab as shown in
. The number
0.1447V
(on the y-axis) can be calculated as a temperature using
:
Temperature = (0.1447 – 0.145300) / 0.00049 + 25 = 23.78°C
A more detailed description of the Scope tab is provided in
.
Figure 33. Eight-Channel Read of Internal Temperature