Using the ADS1298RECG-FE Software
15
SBAU181B – March 2011 – Revised Janurary 2016
Copyright © 2011–2016, Texas Instruments Incorporated
ADS1298R
The
CONTINUOUS
control starts a repeated acquisition process. This function acquires the requested
samples and repeats the data acquisition until the button is turned off. Within a single acquisition cycle,
the points will be contiguous, but from acquisition to acquisition, there may be points missing.
The
Analysis Data input referred
checkbox changes the displayed data that is read from the ADC.
Checking the box displays the data input referred, while not checking displays the data as converted.
The
Show/Poll Lead Off Status
displays a window (see
) that shows the status of the Lead-Off
status registers,
LOFF STATP
and
LOFF STATN
, of the ADS1298R. When the lead for the channel is
disconnected, the corresponding channel LED changes from green to red.
Figure 7. Lead-Off Status Registers Display Window
3.3
About Tab
The
About
tab provides software information to the user. Important safety warning, restrictions, and
disclaimers for the software and hardware are shown and should be followed during the evaluation of this
product. Additional indicators are present to provide device information (
Device ID
and
Rev
) and software
information (
Firmware Version
). The
About
tab should be the first screen displayed at startup (see
).