Using the ADS1298RECG-FE Software
21
SBAU181B – March 2011 – Revised Janurary 2016
Copyright © 2011–2016, Texas Instruments Incorporated
ADS1298R
3.4.4
GPIO and OTHER Registers Tab (ADC Register)
The
GPIO and Other Registers
tab, located under the
Analysis
tab, includes controls for GPIO1 through
GPIO4, respiration phase and frequency, routing of the Wilson amplifiers, and derivation of the Goldberger
terminals.
shows the
GPIO and OTHER Registers
tab and all controls contained on the tab.
Figure 14. GPIO and OTHER Register Tab
The
General-Purpose I/O Register (GPIO)
controls the four general-purpose I/O pins. Each GPIO can be
set as an input or an output via
GPIOCx
controls. If the output is selected, the
GPIODx
control is enabled
allowing the user to set the value to output. If the GPIO is selected as an input, the
GPIODx
control is
disabled and shows the value of the GPIO. If any of the GPIOs are selected as inputs, the
Read GPIO
control is enabled which allows the
GPIODx
values to be updated to the current GPIO value.
The
PACE Detect Register
does not enable a special PACE measurement mode. The register allows for
enabling and configuration of the PACE amplifiers. PACE Amplifier 1 can connect to input channels 1-4
and Pace Amplifier 2 can connect to input channels 5-8.
The
Configuration 4 Register
allows control over the Respiration Frequency, WCT connection to the
RLD and lead-off comparator enable status.
The
Respiration Control Register
allows the user to configure the respiration modulation and
demodulation on Channel 1, Respiration Phase, and set the Respiration Control Signal.