On the static parameter setup interface, there are 24 configurable items.
1.
Pack type: The options include:
a)
Lion: Li-ion battery
b)
NiMH: Nickel-Metal hydride battery
c)
NiCd: Nickel-Cadmium battery
d)
SLA: Maintenance-free lead-acid battery
2.
Pack Vol: The options include:
a)
For Li-ion battery, it may increase by 3.6V per additional battery, and the maximum value
is 14.4V (BTS-2004H).
b)
For Nickel-Metal hydride battery and Nickel-Cadmium battery, it may increase by 1.2V per
additional battery, and the maximum value is 14.4V (BTS-2004H).
c)
For the SLA battery, it may increase by 2V per additional battery, and the maximum value
is 12V (BTS-2004H).
3.
Pack Cap: Set the capacity of the battery. In the subsequent test, the tester will
automatically select the appropriate value for test according to the capacity set here. It may
increase by a step of 50mAH ~ 1000mAH per time, and the maximum value is 10000 mAH
(BTS-2004H).
4.
Charge: Enable or disable the test. Press LEFT or RIGHT to adjust the selection.
5.
Discharge: Enable or disable the test. Press LEFT or RIGHT to adjust the selection.
6.
Over Current: For Li-ion battery, enable or disable the test. For other battery types, this
function is unavailable.
7.
R1: Enable or disable the test. (Identification resistor 1)
8.
R2: Enable or disable the test. (Identification resistor 2)
9.
Vol-Min: It may increase or decrease by a step of 10 mV. When the battery type or battery
voltage is changed, this parameter will be restored to the minimum default value.
10. Vol-Max: It may increase or decrease by a step of 10mV. When the battery type or battery
voltage is changed, this parameter will be restored to the maximum default value.
11. IMP-Min: The maximum value is 0.mΩ.
12. IMP-Max: The maximum value is 999 mΩ. If it is set to be a value larger than 999 mΩ, the
impedance test is disabled.
13. OVC-Min: This parameter can be adjusted only when the over-current test is enabled. It
may increase or decrease at a step of 0.1A.