Entering further basic program data
27
Hysteresis
The respective limit value output only switches after the first, second
or third bad analysis (suppression of the first or the second measured
value). This increases the reliability of the analysis evaluation, e.g.
after the measuring point has been switched over or if the sampling
line has not been flushed sufficiently. The hystereses of the two out-
puts LV1 and LV2 can be set independently of each other.
With a hysteresis of "2", the next analysis is immediately carried out
when the limit value of this analysis has been exceeded. The respec-
tive output is only switched after the limit value of this analysis has
been exceeded for a second time. With a hysteresis of "3", the re-
spective output only switches when the limit value of this analysis has
been exceeded for a third time. This setting is only reactivated once
the value has fallen below the limit value!
(The basic setting for LV1 and LV2 is "1")
In
>BASIC PROGRAM menu
, select
=>
PROGRAM VALUES
=>
HYSTERESIS LV1 or HYSTERESIS LV2
Enter the number of analyses
Press "ENTER" to confirm the entry
Suppression of
bad analyses
ANALYSIS (1,2,3) 1
>
HYSTERESIS LV1
HYSTERESIS LV1
HYSTERESIS LV1
HYSTERESIS LV1
▼▲
ME
ME
ME
ME
ANALYSIS (1,2,3) 1
>
HYSTERESIS LV2
HYSTERESIS LV2
HYSTERESIS LV2
HYSTERESIS LV2
▼▲
ME
ME
ME
ME