Password protection and basic program
24
Minimum interval = 1 litre, maximum interval = 9999 litres. The analy-
sis is carried out once the programmed water quantity has been
measured. The line and the measuring chamber are flushed prior to
the analysis (observe the programmed flush times).
Selecting the quantity control
In
>BASIC PROGRAM menu
, select
=>
PROGRAM VALUES
=>
OPERATING MODE
=>
QUANTITY
INTERVAL
Press "ENTER" to confirm the selection
(An asterisk " * " appears at the end of the line)
The menu
>QUANTITY INTERVAL
appears
Enter the respective flow rate in litres
Press "ENTER" to confirm the entry
Selecting the type of water meter
Select the
>BASIC PROGRAM
menu
=>
PROGRAM VALUES
=>
WATER METER
=>
WATER METER TYPE
Select the water meter constant (litre/impulse)
(100 litres/impulse " * " is factory set)
Press "ENTER" to confirm the selection
Select external analysis triggering
External analysis triggering occurs via a contact at the
start input.
Note: The current analysis interval can be interrupted by triggering a
contact at the
stop input
.
In
>BASIC PROGRAM menu
, select
=>
PROGRAM VALUES
=>
OPERATING MODE
=>
EXTERNAL (START)
Press "ENTER" to confirm the selection
(An asterisk " * " appears at the end of the line)
Quantity control
Start via water meter
External
analysis start
Start via input
000l
>QUANTITY INTERVAL
QUANTITY INTERVAL
QUANTITY INTERVAL
QUANTITY INTERVAL
▼▲
ME
ME
ME
ME
External (Start)
Quantity interval *
TIME CONTROLLED
>
MODE OF OPERATION
MODE OF OPERATION
MODE OF OPERATION
MODE OF OPERATION
▼▲
ME
ME
ME
ME
5 Litres/Impulse
2,5 Litres/Impulse
10 Litres/Impulse
100 Litres/Impulse *
500 Litres/Impulse
1000 Litres/Impulse
1 LITRE/IMPULSE
>
TYPE OF WATER METER
TYPE OF WATER METER
TYPE OF WATER METER
TYPE OF WATER METER
▼▲
ME
ME
ME
ME
External (Start) *
Quantity interval
TIME CONTROLLED
>
MODE OF OPERATION
MODE OF OPERATION
MODE OF OPERATION
MODE OF OPERATION
▼▲
ME
ME
ME
ME