Entering further basic program data
26
Interval pause
If the analysis is triggered via a timer, the interval between two analy-
ses (plus flushing time) is determined by the interval pause. The
shortest interval can be 0 minutes. In this case, analyses are carried
out continuously. The longest interval is 99 minutes.
In
>BASIC PROGRAM menu
, select
=>
PROGRAM VALUES
=>
FLUSH TIMES/INTERVAL
Enter the "
INTERVAL PAUSE"
in minutes (m)
Press "ENTER" to confirm the entry
Limit value monitoring
It is possible to program the limit values on a continuous scale. The
limit value range corresponds to the measuring range of the respec-
tive parameter (water hardness or total chlorine). Two limit value
outputs are available for monitoring. This means that two parameters
(water hardness at measuring point 1, LV1 and total hardness at
measuring point 2, LV2) can be monitored. The functions of the allo-
cated relay outputs can be programmed independently of each other.
The instrument monitors
two measuring points
. Water hardness is
measured at measuring point 1 and total chlorine at measuring point
2. The measuring points are strictly allocated to the limit value out-
puts:
1
LV1 = Measuring point
2
LV2 = Measuring point
2
(water hardness) (total chlorine)
If the limit value LV1 has been exceeded, the limit value control dis-
play
1
lights up RED
and the relay output LV1 reacts as pro-
grammed in the switch function. If the limit value has not been ex-
ceeded, the display lights up GREEN
.
The same applies for the limit
value LV2.
In
>BASIC PROGRAM menu
, select
=>
PROGRAM VALUES
=>
LIMIT VALUES
Enter the values for "
LIMIT VALUE TH"
or
"
LIMIT VALUE CI"
Press "ENTER" to confirm the entry
Monitoring of two
limit values
LIMIT VAL. Cl: 0,5mg/l
LIMIT VAL. TH: 5.0 ppm
>
LIMIT VALUES
LIMIT VALUES
LIMIT VALUES
LIMIT VALUES
▼▲
ME
ME
ME
ME
Interval pause 01m
External flush time
00s
INTERNAL FLUSH TIME
00s
>
FLUSH TIMES/INTERVAL
FLUSH TIMES/INTERVAL
FLUSH TIMES/INTERVAL
FLUSH TIMES/INTERVAL
▼▲
ME
ME
ME
ME