Installation
12
Connecting the power supply and devices
Risk of electric shocks during installation!
If the power supply is not disconnected prior to installation, it may
result in personal injuries, destruction of the product or damage to
plant parts.
Always disconnect the relevant plant parts before installing Tes-
tomat 2000
®
THCL.
Only use tested cables with sufficient cross-sections for the con-
nections.
Risk of damages caused by electromagnetic fields!
If Testomat 2000
®
THCL or the connecting cables are installed
parallel to power cables or in close proximity to electromagnetic
fields, the instrument may be damaged or measurements incor-
rect.
Ensure that connecting cables are as short as possible.
Always install connecting cables and power cables separately.
Connect the instrument to the protective earth conductor (for
230/115 VAC).
Protect Testomat 2000
®
THCL against interference voltages – e.g.
via a mains filter.
Shield the instrument against strong electromagnetic fields.
Block diagram Testomat 2000
®
THCL
Drawn relay positions: Instrument de-energised
WARNING
!
NOTE