5 Parameterizing, adjusting and analyzing software (P2A software)
78
Field Explanation
Measured analog
value
Required field: Entry of the value read off at
the multimeter.
5.3.5.
Transmitter history
Parameterizations, adjustment processes and messages that have
occurred are registered in the transmitter with an operating hours
stamp.
In the history overviews (explained later in more detail), past
processes and events can be made visible.
For parameter changes or adjustments that are performed
directly at the instrument (via the user menu), "Transmitter"
is entered in the
User
field and only the operating hour is
entered in the
Date/time
field instead of operating
hour/date/time.
For entries that are performed using the P2A software, the
name of the user logged into Windows appears in the
User
field, while the date/time and operating hour are shown in
the
Date/time
field.
1.
Mark the instrument file of the connected instrument.
2.
Click on the
[Transmitter history]
button.
- The
Properties of <Instrument type> <Serial number>
dialogue is opened with the
Transmitter history
register.