5 Parameterizing, adjusting and analyzing software (P2A software)
30
Field/button Explanation
Transmitter test
Monitoring of analog outputs
Current reading
Readings are updated every second.
Unit
Unit according to the type of analog
output.
Default value
Freely definable output value for the
respective type of analog output (V or
mA), 1 decimal place.
[Activate]
The entered defined value is
forwarded to the analog output by
clicking. The current reading is frozen.
A warning informs that the value is
being transmitted to the connected
instrument in the event of existing
cabling.
Now check the analog output using a
precise multimeter.
[Deactivate]
Finish entering the electrical variables
at the analog output.
The analog output returns to the
current reading again.
3.
Click on
[OK]
or
[Cancel]
to close the dialogue.
- The analog output returns to Measuring Mode again.