5 Parameterizing, adjusting and analyzing software (P2A software)
33
Field Explanation
How much pressure
is actually applied
Required field: Entry of the value read off at
the pressure sensor.
The n-point adjustment must always be carried out to its full
extent and in good time at all selected adjustment points.
9.
Disconnect connections between the pressure sensor and the
pressure connections of the testo 6321.
Pos: 60 /TD/Produkt verwenden/MUF 63xx/P2A/Analogausgang abgleichen _6321 @ 4\mod_1244461217550_79.doc @ 44819 @ 4
5.3.4.2.
Adjusting the analog output
1.
Connect precision multimeter (see Wiring the instrument page
14).
2.
Mark the instrument file of the connected instrument.
3.
Click on
[Adjusting the transmitter]
.
- The
Properties of <Instrument type> <Serial number>
dialogue is opened with the
Adjusting the transmitter
register.
4.
Click on
[Start wizard …]
and follow the instructions of the
wizard.
-
The adjustment is performed when the wizard is closed.