7 Maintaining the product
75
The condensate outlet must be completely closed
(marking), otherwise measuring errors could be caused by
infiltrated air.
7.9.
Checking / replacing the particle filter
Checking the particle filter:
Check the particle filter of the modular flue gas probe periodically
for contamination: check visually by looking through the window
of the filter chamber.
Replace the filter if there are signs of contamination.
Replacing the particle filter:
The filter chamber may contain condensate.
1.
Open the filter chamber: Turn gently anti-clockwise.
2.
Remove the filter plate and replace it with a new one (0554
3385).
3.
Attach the filter chamber and lock it. Turn gently clockwise.