6 Using the product
48
Performing recalibration / adjustment:
WARNING
Dangerous gases
Danger of poisoning!
> Observe safety regulations / accident prevention regulations
when handling test gas.
> Use test gases in well ventilated rooms only.
1.
Connect the calibration adapter to the flue gas socket.
2.
Select the parameter:
[
▲
]
,
[
▼
]
→
[OK]
.
3.
[Edit]
→
Enter the test gas concentration (nominal value).
4.
Attach the connecting line of the test gas bottle to the
calibration adapter.
5.
Apply test gas to the sensor.
6.
Start recalibration:
[Start]
.
7. Accept the nominal value once the actual value is stable
(adjustment):
[OK]
.
-or-
Abort (no adjustment):
[esc]
.
8. Save changes:
[Finished]
.
6.1.4.
Fuels
The fuel can be selected. The fuel-specific coefficients and limits
can be set.
Apart from the pre-configured fuels, 10 more customer specific
fuels can be configured. Fuel parameter, see
www.testo.com/download-center (registration required).
In order to maintain the measuring accuracy of the
instrument one must choose or configure the correct fuel.
Correct representation of measuring results is only assured
if the threshold values for the ideal range of the
corresponding measurement task have been set correctly.
The pre-set threshold values are typical values for the
selected system type and the chosen type of fuel.
Calling up the function:
>
[ ]
→
Fuels
→
[OK]
.