7.5 Checking the distance between mixing bowl and stirrer
If the distance between the mixing bowl and the stirrer is greater or lesser than the ap-
plicable standard allows for, this can be adjusted.
First, the mixing bowl must be lowered and removed (see point
Now remove the contact protection. The contact protection is fixed with 3 spring ball
pressure pieces, it only has to be pressed down.
Figure 5 Removing contact protection
Reinsert the mixing bowl and stirrer.
The stirrer can now be turned by hand.
With the spacing template (TESTING Art.-No.: 1.0203.07), the distance can be checked
and adjusted as described in point
(see: Height adjustment). If the distance con-
forms to the standards, the contact protection is refitted.
Warning
Disconnect the power plug before starting work!
This could result in serious injuries, as your hands can get caught in
the
rotating stirrer.
Warning
Warning - do not use the mixer without contact protection!
This could result in serious injuries, as your hands can get caught in
the
rotating stirrer.
The contact protection en-
gages at these points.