Technical data
24
Testboy
®
TV 225
Technical data
Test procedure
True RMS
Accuracy
DC voltage
± 0.8 % + 3 digits, min. 0.2 V
Input impedance 2 MΩ
AC voltage
± 1.2 % + 3 digits, min. 0.5 V at 45 – 65 Hz
± 2.0 % + 5 digits, min. 0.5 V at 65 – 2000 Hz
Input impedance 2 MΩ
Alternating current
± 3.0 % + 5 digits
Frequency
± 0.5 % + 2 digits
Resistance
± 1.0 % + 3 digits
Continuity
Beep < 50 Ω, test current 1 mA
Scan frequency
3 Hz
Operating conditions
0 – 40 °C at < 80 % RH
< 2000 m above MSL
Dimensions LxWxH
324 x 178 x 30 mm
Protection class
IP 42
Overvoltage category
CAT IV 600 V
Testing standard
IEC/EN 61010-1, 61010-2-032, 61326-1
Voltage supply
3x 1.5 V type AAA Micro