background image

6

 - [EN]  English

K345 Issue 4

Operation

This section gives examples of how to connect and use the 
instrument. Before you start:
• Read and understand the “Safety” section.
• Do not use a damaged instrument.

Operation - RTD connections

To prevent instrument errors, make sure that the RTD 
connections (A1-item [10]) are correct. In the examples 
that follow 2W, 3W, and 4W identify the 2, 3, and 4 wire 
connections.

Operation - Communications port connections

Use the communications port (A1 - item [9]) to attach an 
IDOS Universal Measurement Module (UMM).
When you attach the cable from a UMM (Figure 7/8), the 
instrument automatically changes the menus to give you 
all the applicable options (Table 2/3).

Operation - Measure RTD values

To measure the temperature or resistance values of an 
RTD:
1. Connect the instrument (Figure 1) and, if necessary, 

adjust the 

Set Up

 (Table 1).

2. Select the RTD or Ohms input task from 

Select Task

 

(Table 2/3) and, if necessary, adjust the 

Settings

 

(Table 4)

Figure 1: 

Example configuration - To measure the 

temperature or resistance of an RTD

The display shows the number of RTD connections.

If this symbol does not agree with the number of RTD 
connections: 
• Make sure that the RTD connections are correct.
• Make sure that the wires and the sensor are 

serviceable.

Operation - Change the output values

When the output operation is set up (Table 5), use one of 
these procedures to change the output values:

Table 7: 

Procedures to change the output 

Example procedure (“Nudge” output):

a) Temperature

b) Resistance

Settings

Settings

°C

Pt100

 = Four-wire RTD attached.

Output

Procedure

Select 

Edit

 (

 

) and/or use the 

 

 buttons. 

See the example below.

Select 

Start/Stop

 (

 

) or use the 

 

 buttons 

to make the step changes manually. 
Select 

Start/Stop

 (

 

). 

Display: 

RTD output

1

Edit

2

3

Edit

4

 

 

■ ▲

+100.00

Sk1 = Edit
Sk2 = Settings

Display: 

RTD output

5

6

101.00

102.00

Increment = 1 (Table 5

Summary of Contents for Druck DPI 811

Page 1: ...Druck DPI 811 812 RTD calibrator and RTD loop calibrator User manual K345 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Page 2: ...K345 Issue 4 A1 A2 DPI 812 A3 10 1 3 4 2 5 6 7 9 8 A2 A3 11 B1 12 10 14 15 13 20 21 22 25 17 18 26 19 27 16 23 24 B1 Customer service Visit our web site www gesensing com...

Page 3: ...s uses Intelligent Digital Output Sensor IDOS technology to give instant plug and play functionality with a range of Universal Measurement Modules UMM Example the Universal Pressure Module UPM The DPI...

Page 4: ...ion above it on the display Item 25 Example Settings 8 Display Refer to A3 9 SENSOR PC Communications port Use to connect a Universal Measurement Module UMM or a RS232 cable 10 RTD connectors To measu...

Page 5: ...se this instrument together with a HART communicator to set up and calibrate HART devices Scale To select the applicable international temperature scale IPTS 68 or ITS 90 To select and set up the back...

Page 6: ...select and set up a tare value for the measurement task a specified value or the reading on the display The instrument subtracts a positive tare value and adds a negative tare value Additional data S...

Page 7: ...ional Additional data Table 6 Select Settings Step To select and set up values for the Defined Step output Example output cycle Auto Repeat Optional Additional data Table 6 Select Settings Options If...

Page 8: ...adjust the Set Up Table 1 2 Select the RTD or Ohms input task from Select Task Table 2 3 and if necessary adjust the Settings Table 4 Figure 1 Example configuration To measure the temperature or resis...

Page 9: ...ransmitter calibration DPI 812 only To calibrate a transmitter 1 Connect the instrument Figure 3 4 and if necessary adjust the Set Up Table 1 2 Select the applicable calibration task from Select Task...

Page 10: ...the top right hand corner 3 Supply the output values to the system Table 7 Example Nudge output a Use Edit to set a value less than the switch value b Use the buttons to change the value in small incr...

Page 11: ...Settings Table 4 5 2 If necessary do a zero correction Table 4 To measure pressure with another operation Figure 8 use the same procedure Figure 8 Example configuration To measure pressure and temper...

Page 12: ...set the new calibration date for the instrument Calibration Procedures Ohms input 1 Let the equipment get to a stable temperature minimum 5 minutes since the last power on 2 Use the calibration menu...

Page 13: ...nt get to a stable temperature minimum 5 minutes since the last power on 3 Use the calibration menu Table 9 to do a three point calibration FS Zero and FS The display shows the applicable instructions...

Page 14: ...850 0 90 0 50 Pt100 385 IEC 751 328 1562 200 850 0 45 0 25 Pt200 385 IEC 751 328 1562 200 850 1 08 0 60 Pt500 385 IEC 751 328 1562 200 850 0 72 0 40 Pt1000 385 IEC 751 328 752 200 400 0 36 0 20 D 100...

Reviews: