background image

10

 - [EN]  English

K345 Issue 4

Maintenance

This section gives procedures to maintain the unit in a 
good condition. Return the instrument to the 
manufacturer or an approved service agent for all repairs.

 

Do not dispose of this product as household waste. Use an 
approved organisation that collects and/or recycles waste 
electrical and electronic equipment.

Maintenance - Clean the unit

Clean the case with a moist, lint-free cloth and a weak 
detergent. Do not use solvents or abrasive materials.

Maintenance - Replace the batteries 

 

 

To replace the batteries, refer to B1. Then re-attach the 
cover.
Make sure that the time and date are correct. The 
calibration facility uses the date to give service and 
calibration messages.
All the other configuration options stay in memory.

Calibration

Note: GE can provide a calibration service that is traceable 
to international standards.

We recommend that you return the instrument to the 
manufacturer or an approved service agent for 
calibration.
If you use an alternative calibration facility, make sure that 
it uses these standards.

Calibration - Before you start

To do an accurate calibration, you must have:
• the calibration equipment specified in Table 8.
• a stable temperature environment: 70 ± 2°F (21 ± 1°C)

Table 8: 

Calibration equipment

Before you start the calibration, make sure that the time 
and date on the instrument are correct (Table 1).

Selection sequence: 

 Select Task 

(Table 2)

 

 Set Up 

(Table 1)

 

 Calibration 

.

Table 9: 

Calibration options

When you select a channel, the display shows the 
applicable instructions to complete the calibration.
When the calibration is complete, select 

Calibration Due 

and set the new calibration date for the instrument.

Calibration - Procedures (Ohms input)

1. Let the equipment get to a stable temperature 

(minimum: 5 minutes since the last power on).

2. Use the calibration menu (Table 9) to do a two-point 

calibration for each range. 

• Range: 0-399.9

Ω

a. Nominal zero ohms: Make a 4 wire connection to 

the 0

Ω

 resistor (Figure 1).

b. Nominal positive full-scale ohms: Make a 4 wire 

connection to the 400

Ω

 resistor (Figure 1)

• Range: 400

Ω

-4k

Ω

a. Nominal zero ohms: Make a 4 wire connection to 

the 400

Ω

 resistor (Figure 1).

b. Nominal positive full-scale ohms: Make a 4 wire 

connection to the 4k

Ω

 resistor (Figure 1)

The display shows the applicable instructions to 
calibrate each range.

Function

Calibration equipment

(ppm = parts per million)

 Ohms

- Standard 0

Ω

 resistor

- *Standard resistor (

Ω

): 100, 200, 300

Tolerance: 50 ppm + 0.6 ppm/°C + 5 ppm/year

*

Or an equivalent resistance simulator

- *Standard resistor (

Ω

): 400, 1k, 2k, 4k

Tolerance: 10 ppm + 0.6 ppm/°C + 5 ppm/year

 

Ohms

An ohmmeter or an RTD measurement system 
with these limits for the excitation current:

 range (

Ω

):

Excitation (mA)

0 ...   400

Ω

:  0.50 to 3.0 mA

400 ... 1500

Ω

:  0.05 to 0.8 mA

1500 ... 3200

Ω

: 0.05 to 0.4 mA

3200 ... 4000

Ω

: 0.05 to 0.3 mA

Pressure

UPM only. Refer to the user manual for the IDOS 
UPM.

mA

mA calibrator. 
Accuracy: Refer to Table 12.

B1

1

Display: Enter 

Calibration PIN

2

3

Menu:

Select Channel

4

5

+

 

Factory PIN = 4321

(Table 9)

Options

Description

Ohms To calibrate the input resistance.
Ohms To calibrate the output resistance

 IDOS ...

UMM only. To calibrate the specified IDOS UMM. 
Refer to the user manual for the IDOS UMM.

mA

DPI 812 only. To calibrate the mA input.

Calibration Due: 

To set the date of the next 

calibration for the instrument.
After the specified calibration date, there is a 
warning message. There is a selection box to stop 
the warning.
To change the calibration PIN (Personal 
Identification Number).

Summary of Contents for Druck DPI 811

Page 1: ...Druck DPI 811 812 RTD calibrator and RTD loop calibrator User manual K345 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Page 2: ...K345 Issue 4 A1 A2 DPI 812 A3 10 1 3 4 2 5 6 7 9 8 A2 A3 11 B1 12 10 14 15 13 20 21 22 25 17 18 26 19 27 16 23 24 B1 Customer service Visit our web site www gesensing com...

Page 3: ...s uses Intelligent Digital Output Sensor IDOS technology to give instant plug and play functionality with a range of Universal Measurement Modules UMM Example the Universal Pressure Module UPM The DPI...

Page 4: ...ion above it on the display Item 25 Example Settings 8 Display Refer to A3 9 SENSOR PC Communications port Use to connect a Universal Measurement Module UMM or a RS232 cable 10 RTD connectors To measu...

Page 5: ...se this instrument together with a HART communicator to set up and calibrate HART devices Scale To select the applicable international temperature scale IPTS 68 or ITS 90 To select and set up the back...

Page 6: ...select and set up a tare value for the measurement task a specified value or the reading on the display The instrument subtracts a positive tare value and adds a negative tare value Additional data S...

Page 7: ...ional Additional data Table 6 Select Settings Step To select and set up values for the Defined Step output Example output cycle Auto Repeat Optional Additional data Table 6 Select Settings Options If...

Page 8: ...adjust the Set Up Table 1 2 Select the RTD or Ohms input task from Select Task Table 2 3 and if necessary adjust the Settings Table 4 Figure 1 Example configuration To measure the temperature or resis...

Page 9: ...ransmitter calibration DPI 812 only To calibrate a transmitter 1 Connect the instrument Figure 3 4 and if necessary adjust the Set Up Table 1 2 Select the applicable calibration task from Select Task...

Page 10: ...the top right hand corner 3 Supply the output values to the system Table 7 Example Nudge output a Use Edit to set a value less than the switch value b Use the buttons to change the value in small incr...

Page 11: ...Settings Table 4 5 2 If necessary do a zero correction Table 4 To measure pressure with another operation Figure 8 use the same procedure Figure 8 Example configuration To measure pressure and temper...

Page 12: ...set the new calibration date for the instrument Calibration Procedures Ohms input 1 Let the equipment get to a stable temperature minimum 5 minutes since the last power on 2 Use the calibration menu...

Page 13: ...nt get to a stable temperature minimum 5 minutes since the last power on 3 Use the calibration menu Table 9 to do a three point calibration FS Zero and FS The display shows the applicable instructions...

Page 14: ...850 0 90 0 50 Pt100 385 IEC 751 328 1562 200 850 0 45 0 25 Pt200 385 IEC 751 328 1562 200 850 1 08 0 60 Pt500 385 IEC 751 328 1562 200 850 0 72 0 40 Pt1000 385 IEC 751 328 752 200 400 0 36 0 20 D 100...

Reviews: