4

 - [EN]  English

K345 Issue 4

To start - Set up the settings

When the task is set up (Table 2/3), use the 

Settings

 menu 

to adjust the input and/or output operation

.

If there is additional data for a menu option, select 

Settings

 (

 

) to see the values that are set up. If 

necessary, adjust the values. Refer to “Edit functions”.

Table 4: 

(

Part of table) Menu options - Settings (Input)

Table 4: 

(Part of table) Menu options - Settings (Input) 

To start - Edit functions

Example 1) Set up a label for 

x:y Scaling

 = %.

Display: 

RTD + mA

1

Settings selection 

(If applicable)

2

3

 

Sk1 = Start/Stop
Sk2 = Settings

Menu:

Settings

4A

5A

6A

Settings

x:y 

 

(Table 4)

[

]/[  ]

Sk1 = Edit

Menu:

Settings

4B

5B

(Table 5)

[

]/[  ]

Options

(If applicable)

Description

... Units To select the temperature units (°C or °F).

UPM only = “Pressure Units” if you select an IDOS 
task (Table 2/3
). Select one of the fixed units of 
measurement (psi, mbar ... ). 

... type To select an applicable RTD type (Pt50, Pt100 ... ) 

To include maximum, minimum and average 
values for the measurement task. 
To select and set up a tare value for the 
measurement task (a specified value or the reading 
on the display). 
The instrument subtracts a positive tare value, and 
adds a negative tare value.

Additional data: Select Settings 

(

 

)

Options

(If applicable)

Description

To select and set up a scale of values: One local 
scale for each measurement task (Maximum: 5). 

Additional data (Example 1/2): Select Settings 

(

 

)

To select and set up the filter values to give a 
smoother output for the measurement task:

Band as a % of full scale (FS). The filter 
compares each new value with the previous 

value. If the new value is outside the band, it is not 
filtered.

Low pass filter time constant in seconds. 
Increase the value to increase damping 

factor.

Additional data: Select Settings 

(

 

)

To select and set up the alarm values for the 
measurement task (maximum and minimum).

Additional data: Select Settings 

(

 

)

UPM only.

 Gage sensors or sensors with differential 

operation. A zero correction that makes the 
instrument read zero at local pressure.
Leak Test only. To set an applicable period for the 
leak test (Hours:Minutes:Seconds).

Settings

x:y 

1

Edit

 

2

Edit

3

 

 

Sk1 = Edit

A3 - item [27]

Edit

 

...

Edit

4

5

%_

 

A3 - item [27]

Shift

OK

Summary of Contents for Druck DPI 811

Page 1: ...Druck DPI 811 812 RTD calibrator and RTD loop calibrator User manual K345 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Page 2: ...K345 Issue 4 A1 A2 DPI 812 A3 10 1 3 4 2 5 6 7 9 8 A2 A3 11 B1 12 10 14 15 13 20 21 22 25 17 18 26 19 27 16 23 24 B1 Customer service Visit our web site www gesensing com...

Page 3: ...s uses Intelligent Digital Output Sensor IDOS technology to give instant plug and play functionality with a range of Universal Measurement Modules UMM Example the Universal Pressure Module UPM The DPI...

Page 4: ...ion above it on the display Item 25 Example Settings 8 Display Refer to A3 9 SENSOR PC Communications port Use to connect a Universal Measurement Module UMM or a RS232 cable 10 RTD connectors To measu...

Page 5: ...se this instrument together with a HART communicator to set up and calibrate HART devices Scale To select the applicable international temperature scale IPTS 68 or ITS 90 To select and set up the back...

Page 6: ...select and set up a tare value for the measurement task a specified value or the reading on the display The instrument subtracts a positive tare value and adds a negative tare value Additional data S...

Page 7: ...ional Additional data Table 6 Select Settings Step To select and set up values for the Defined Step output Example output cycle Auto Repeat Optional Additional data Table 6 Select Settings Options If...

Page 8: ...adjust the Set Up Table 1 2 Select the RTD or Ohms input task from Select Task Table 2 3 and if necessary adjust the Settings Table 4 Figure 1 Example configuration To measure the temperature or resis...

Page 9: ...ransmitter calibration DPI 812 only To calibrate a transmitter 1 Connect the instrument Figure 3 4 and if necessary adjust the Set Up Table 1 2 Select the applicable calibration task from Select Task...

Page 10: ...the top right hand corner 3 Supply the output values to the system Table 7 Example Nudge output a Use Edit to set a value less than the switch value b Use the buttons to change the value in small incr...

Page 11: ...Settings Table 4 5 2 If necessary do a zero correction Table 4 To measure pressure with another operation Figure 8 use the same procedure Figure 8 Example configuration To measure pressure and temper...

Page 12: ...set the new calibration date for the instrument Calibration Procedures Ohms input 1 Let the equipment get to a stable temperature minimum 5 minutes since the last power on 2 Use the calibration menu...

Page 13: ...nt get to a stable temperature minimum 5 minutes since the last power on 3 Use the calibration menu Table 9 to do a three point calibration FS Zero and FS The display shows the applicable instructions...

Page 14: ...850 0 90 0 50 Pt100 385 IEC 751 328 1562 200 850 0 45 0 25 Pt200 385 IEC 751 328 1562 200 850 1 08 0 60 Pt500 385 IEC 751 328 1562 200 850 0 72 0 40 Pt1000 385 IEC 751 328 752 200 400 0 36 0 20 D 100...

Reviews: