background image

2

 - [EN]  English

K345 Issue 4

Safety (Continued)

• UPM only. To prevent a dangerous release of 

pressure, isolate and bleed the system before you 
disconnect a pressure connection.

Before you start an operation or procedure in this 
publication, make sure that you have the necessary skills 
(if necessary, with qualifications from an approved training 
establishment). Follow good engineering practice at all 
times.

Safety - Marks and symbols on the instrument

To start

To start - Location of items 

 

 …

 

To start - Items on the display 

 

To start - Prepare the instrument

Before you use the instrument for the first time:
• Make sure that there is no damage to the instrument, 

and that there are no missing items.

• Remove the plastic film that protects the display. Use 

the tag (

 ) in the top right-hand corner.

• Install the batteries (refer to B1). Then re-attach the 

cover.

Complies with European 
Union directives

Warning - refer to the 
manual

Read the manual

Battery

Ground (Earth) 

 

ON/OFF

Do not dispose of this product as household waste. Refer to 
“Maintenance”.

Item

Description

1.

On or off button. 

2.

 

Left-hand soft-key. Selects the function above it on 
the display (Item 25). Example: Edit

3.

ESC

Moves back one menu level.
Leaves a menu option.
Cancels the changes to a value.

4.

Increases or decreases a value.
Highlights a different item.

5.

HOLD

Holds the data on the display. To continue, press the 

HOLD

 button again. 

6.

MENU

OK

Shows the 

Select Task

 menu.

Selects or accepts an item or value.
Selects [

] or cancels [  ] a selection.

7.

 

Right-hand soft-key. Selects the function above it on 
the display (Item 25). Example: Settings

8.

Display. Refer to A3

9.

SENSOR

/ PC

Communications port. Use to connect a Universal 
Measurement Module (UMM) or a RS232 cable.

10.

RTD connectors: To measure, use the 2 Wire, 3 Wire, 
or 4 Wire “INPUT” connectors. To simulate, use only 
the two “OUT” connectors. Refer to “Operation”.

11.

Connection point for some of the optional 
accessories. Refer to the datasheet.

12.

Battery compartment. Refer to B1.

13., 14., 15.

DPI 812 only. Terminals to measure current, to supply 
24V source, and to do switch tests.

A1

A2

Item

Description

16.

DPI 812 only. Task indication for the switch test. 

= switch closed

 = switch open

UPM only. Task indication for the leak test.

Refer to: Select Task (Table 2/3)

17.

DPI 812 only. The loop power supply is on.

Refer to: Select Task (Table 2/3)

18.

The measured value satisfies one of the alarm 
conditions. 

Refer to: Settings (

Table 4

)

19.

The data on the display is on hold. To continue, press 
the 

HOLD

 button again. 

20.

Shows the battery level: 0 ... 100%.

21.

Identifies the type of data and the measurement 
range.

= Input

= Output

= IDOS input

Refer to: Select Task (Table 2/3)

22. ... 24.

The settings applied to the input or output.

22.

°C

The units or a specified scale (x:y) - (Table 4/

5

).

Pt...

The RTD type (Pt50, ...) - (Table 4/

5

). 

RTD input connections: 2, 3, or 4 (Figure 1)

23.

, ... ,

 

= Output operation (Table 5)

24.

= Filter

= Tare

= Maximum
= Average
= Minimum

(Table 4)

25.

A soft-key function. To select an available function, 
press the soft-key below it. Example:

= Move left 

= Move right

26.

The measured value or values applicable to the task 
selection.

27.

The 

Edit

 display to set up text labels ( 

 6 characters): 

x:y

 

Scaling (

Table 4

)

.

OK

= Accept the new text label

Shift

= Change the keys: 123ABC or -_+abc
= Add a space

BS

= Back space (Delete character)

A3

Summary of Contents for Druck DPI 811

Page 1: ...Druck DPI 811 812 RTD calibrator and RTD loop calibrator User manual K345 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Page 2: ...K345 Issue 4 A1 A2 DPI 812 A3 10 1 3 4 2 5 6 7 9 8 A2 A3 11 B1 12 10 14 15 13 20 21 22 25 17 18 26 19 27 16 23 24 B1 Customer service Visit our web site www gesensing com...

Page 3: ...s uses Intelligent Digital Output Sensor IDOS technology to give instant plug and play functionality with a range of Universal Measurement Modules UMM Example the Universal Pressure Module UPM The DPI...

Page 4: ...ion above it on the display Item 25 Example Settings 8 Display Refer to A3 9 SENSOR PC Communications port Use to connect a Universal Measurement Module UMM or a RS232 cable 10 RTD connectors To measu...

Page 5: ...se this instrument together with a HART communicator to set up and calibrate HART devices Scale To select the applicable international temperature scale IPTS 68 or ITS 90 To select and set up the back...

Page 6: ...select and set up a tare value for the measurement task a specified value or the reading on the display The instrument subtracts a positive tare value and adds a negative tare value Additional data S...

Page 7: ...ional Additional data Table 6 Select Settings Step To select and set up values for the Defined Step output Example output cycle Auto Repeat Optional Additional data Table 6 Select Settings Options If...

Page 8: ...adjust the Set Up Table 1 2 Select the RTD or Ohms input task from Select Task Table 2 3 and if necessary adjust the Settings Table 4 Figure 1 Example configuration To measure the temperature or resis...

Page 9: ...ransmitter calibration DPI 812 only To calibrate a transmitter 1 Connect the instrument Figure 3 4 and if necessary adjust the Set Up Table 1 2 Select the applicable calibration task from Select Task...

Page 10: ...the top right hand corner 3 Supply the output values to the system Table 7 Example Nudge output a Use Edit to set a value less than the switch value b Use the buttons to change the value in small incr...

Page 11: ...Settings Table 4 5 2 If necessary do a zero correction Table 4 To measure pressure with another operation Figure 8 use the same procedure Figure 8 Example configuration To measure pressure and temper...

Page 12: ...set the new calibration date for the instrument Calibration Procedures Ohms input 1 Let the equipment get to a stable temperature minimum 5 minutes since the last power on 2 Use the calibration menu...

Page 13: ...nt get to a stable temperature minimum 5 minutes since the last power on 3 Use the calibration menu Table 9 to do a three point calibration FS Zero and FS The display shows the applicable instructions...

Page 14: ...850 0 90 0 50 Pt100 385 IEC 751 328 1562 200 850 0 45 0 25 Pt200 385 IEC 751 328 1562 200 850 1 08 0 60 Pt500 385 IEC 751 328 1562 200 850 0 72 0 40 Pt1000 385 IEC 751 328 752 200 400 0 36 0 20 D 100...

Reviews: