RS-232C Commands
45
Type
Command
Input
FAIL
Response
{ OK | ERR40 }
●
OK : turns the red FAIL LED on.
●
ERR40: Status Error. The command is sent while the lid is open.
5.5.7 Error Data List
Table 5-2
The List of Error Data
Error Code
Description
ERR10
Syntax error. The command does not exist or is invalid.
ERR20
It only occurs when the Handler is in Sensor Mode. Inward Handler Sensor
detection error. (The horizontal operation of fixture is incomplete.)
ERR30
It only occurs when the Probe is in Sensor Mode. Downward Probe Sensor
detection error. (The vertical operation of fixture is incomplete.)
ERR40
Status Error, It occurs when an invalid command is sent. For example, a Handler
OUTWARD or Probe UPWARD while the lid is closed. It is returned if the command
is sent when the motion can not be done in conjunction with the current status.
ERR50
Lid Open Sensor error. It occurs only to certain products which should always
detect whether the lid is open. The response comes when the sensor does not
detect open lid.
ERR60
Safety error. It is returned when a problem in multiple motions occurs.
ERR90
It occurs when the command is sent while the Handler or the Probe is in Off Mode.
Summary of Contents for TC-5916AP
Page 1: ...TC 5916AP Pneumatic Shield Box User Manual R20190930...
Page 5: ...Revision History v Revision Record Revision Description Date v1 0 Initial Release 2019 09 23...
Page 9: ...PART 1 Getting Started 1...
Page 13: ...PART 2 Hardware Description 5...
Page 18: ...10 Dimension 2 2 Dimension...
Page 19: ...Dimension 11...
Page 25: ...PART 3 Putting into Operation 17...
Page 55: ...PART 4 Maintenance 47...
Page 63: ...Expendables and Accessories 55 Figure 7 1 TC 5916AP Expendables Location...