RS-232C Commands
43
5.5.5 Probe Operation Commands
PROBE?
UPWARD
DOWNWARD
PROBE?
A query command to check the current status of the Probe.
Type
Query
Input
PROBE?
Response
{ UPWARD | DOWNWARD }
●
UPWARD: The Probe is at upward position.
●
DOWNWARD: The Probe is at downward position.
UPWARD
A command to lift the Probe up.
Type
Command
Input
UPWARD
Response
{ OK | ERR40 | ERR90 }
●
OK: The Probe is at upward position.
●
ERR40: Status Error, It occurs when an invalid command is sent. For example:
○
Handler OUTWARD or Probe UPWARD command is sent while the lid is closed.
○
Probe DOWNWARD or UPWARD command is sent while the Handler is at outward
position.
○
Handler OUTWARD command is sent while the Probe is at downward position.
●
ERR90: The command is sent while the Probe is in Off Mode.