POWER PANEL
plus
M
N° ME07.99077
67
B. Vuille
Probing gauge block
X1
Probe first point
on the gauge
block.
X2
Probe second
point on the
gauge block.
Relevant angle =
arc tg (P2-P1)/X
is being
calculated.
Display shows
calculated value
in selected unit.
Conversion into degree, minute and second
Conversion into decimal degree
Conversion into radian
Storage of X value for all subsequent angle
measurements. If a part programme is created by learning,
this value remains constantly available in
the memory and it will not be necessary to enter it again bye
executing the programme.
Perform new
probe contact
or exit Mode 1