18) Executing a part programme
The configuration menu provides a number of parameters which may need to be
specifically set for each part programme. The settings that need be made for running
this part programme are described as follows:
-
Probing speed: 7,5 / 15 mm/s
-
Fast displacement: 20 / 30 / 40 mm/s. Probe will automatically be positioned at
the selected speed when starting probing.
-
Response time of display: given value indicates in seconds how long display
remains blocked before next numbered value appears. This options is enabled
when AUTO is set to ON.
-
Waiting time for stabilisation: given value indicates in seconds how long you’ve
to wait before starting probing after the probe has reached its correct position.
This option is enabled when POS is set to ON.
AUTO
POS
Prompt
for next
dimension
Positioning
Response
time of
display
Waiting
time for
stabilisation
Displayed
value
OFF
OFF
Enter
Manual
+ Enter
to start
probing
Infinite
Infinite
Last measured
value.
Next value is
prompted once
Enter is pressed.
OFF
ON
Enter
Auto
Infinite
As set
Last measured
value.
Probe is positioned
once Enter is
pressed.
Next value is
prompted once
given waiting time
is over.
ON
OFF
Auto
Manual
+ Enter
to start
probing
As set
Infinite
Last measured
value.
Next value is
prompted once
given response
time is over.
Subsequent value
is prompted once
Enter is pressed.
ON
On
Auto
Auto
As set
As set
Last measured
value.
Probe is positioned
once response
time is over.
Nex value is
prompted once
waiting time is
over.
Note :
The function key F6 may be activated instead of Enter or the power
control may be rotated by one click.