POWER PANEL
plus
M
N° ME07.99077
84
B. Vuille
12) PAUSE
Use this function key to activate the still mode. Once this is done, you may continue
measuring based on a new reference C. The measured values will not be kept in the
memory and the measurement cycle will not be interrupted. When capturing
deviations in parallelism and perpendicularity, this function also serves to temporarily
stop value capture, e.g. when getting over a groove (see chapter 9 and 11 for more
information).
PAUSE can be
prompted from
the measuring
mode.
System has
switched over
to the standby
condition.
To access reference C temporarily
or
To exit PAUSE mode and return to previous stage.
In the PAUSE mode, you’re not allowed to capture a new probe constant
as this operation could affect the measurement cycle being carried out at
the time this mode was enabled. You may, however, use the functions
enabling you to perform all needed probe contacts or measure distances
or medians.