The name of the table can be edited so that it is easier to distinguish from other tables.
The table’s name cannot be longer than 11 characters. You can also edit the percentage
and SI reading columns in the table. In the example below,
Point 1
is for 1 percent and
has the assigned value of 26.0 X 10
3
SI units.
Figure 135: Concentration Table
When creating a table, you do not need to have all of the 12 data points entered for
reference. Smaller tables, with a minimum of 4 data points, can be used, but they must
follow a specific format.
In order to use a smaller table, the concentration table must be padded with zeroes where
there are unpopulated data points. The absolute position of the table is not important, but
it cannot contain zeroes between the start and end data points as this will result in the
meter neglecting the higher concentration data points
(Figure 136)
.