Service Information

Preparing for Calibration

5

Table 4.  Performance Test Values

Step

Calibrator
Setting

50T Coil
Current

Measure

 i410
Lo mV

i410
Hi mV

 i1010
Lo mV

i1010
Hi mV

DC Amps Zero Test

1

0.0 

(ZERO knob CW)

NA

Pos. span

3.0

15.0

3.0

15.0

2

0.0 

(ZERO knob CCW)

NA

Neg. span

15.0

3.0

15.0

3.0

DC Amps Test

1

3

0.04A dc

2.0A dc

2.00 mV

1.43

2.57

1.46

2.54

4

0.4A dc

20.0A dc

20.0 mV

18.8

21.2

19.1

20.9

5

-0.4A dc

-20.0A dc

-20.0 mV

-21.2

-18.8

-20.9

-19.1

6

1.2A dc

60.0A dc

60.0 mV

57.4

62.6

58.3

61.7

AC Amps Test

7

0.40A, 45 Hz, 400 Hz

20.0A ac

20.0 mV

18.8

21.2

19.1

20.9

8

0.40A, 2 kHz

20.0A ac

20.0 mV

NA

NA

18.9

21.1

9

1.0A 45 Hz, 400 Hz

50.0A ac

 50.0 mV

47.7

52.3

48.5

51.5

10

1.0A 2 kHz

50.0A ac

 50.0 mV

NA

NA

48.0

52.0

11

2.0A, 45 Hz, 400 Hz

100.0A ac

100.0 mV

96.0

104.0

97.5

102.5

12

2.0A, 1 kHz

100.0A ac

100.0 mV

NA

NA

96.5

103.5

DC Amps Balance Test

13

4.0A dc

200.0A dc

200.0 mV

1.0A maximum change from Center of Jaw
Reading to Top or Bottom of Jaw Readings.

DC Amps Test

14

4.0A dc

200.0A dc

200.0 mV

192.5

207.5

195.5

204.5

15

8.0A dc

400.0A dc

0.4000V

0.3855V

0.4145V

0.3915V

0.4085V

16

-8.0A dc

-400.0A dc

-0.4000V

-0.4145V

-0.3855V

NA

NA

17

-12.0A dc

-600.0A dc

-0.6000V

NA

NA

-0.6125V

-0.5875V

18

20.0A dc

1000.0A dc

1.0000V

NA

NA

0.9795V

1.0205V

19

-20.0A dc

-1000.0A dc

-1.0000V

NA

NA

-1.0205V

-0.9795V

AC Amps Test

20

2.4A, 45 Hz, 400 Hz

120.0A ac

120.0 mV

115.3

124.7

117.1

122.9

21

2.4A, 800 Hz

120.0A ac

120.0 mV

NA

NA

115.9

124.1

22

5.0A, 45 Hz, 400 Hz

250.0A ac

250.0 mV

240.7

259.3

244.5

255.5

23

8.0A, 45 Hz, 400 Hz

400.0A ac

0.4000V

0.3855V

0.4145V

NA

NA

24

12.0A, 45 Hz, 65 Hz

600.0A ac

0.6000V

NA

NA

0.5875V

0.6125V

Notes:

1. Prior to performing DC Amp tests, adjust ZERO so that the DMM reads 00.0 +/- 00.1 mV without current applied.

Summary of Contents for FLUKE i1010

Page 1: ...p use only specified replacement parts Caution If the Current Clamp is to be Calibrated you must have a replacement decal See the Replacement Parts section for further details This Current Clamp conta...

Page 2: ...Product 240 000 Storage Temperature 20 to 60 C 4 to 140 F Output Signal 1 mV per amp dc or ac Relative Humidity 0 to 95 0 to 30 C 0 to 75 30 to 40 C 0 to 45 40 to 50 C Working Voltage 600V CAT III ma...

Page 3: ...n Battery LED Test Complete the following procedure to verify that the Battery LED threshold is set correctly 1 Turn on the Calibrator and allow it to warm up and stabilize for 30 minutes 2 Place the...

Page 4: ...ollowing 1 Calibration necessitates removal of the decal on the case top Plan to install a new decal when cali bration is complete See Replacement Parts for decal ordering information 2 Turn on the Ca...

Page 5: ...104 0 97 5 102 5 12 2 0A 1 kHz 100 0A ac 100 0 mV NA NA 96 5 103 5 DC Amps Balance Test 13 4 0A dc 200 0A dc 200 0 mV 1 0A maximum change from Center of Jaw Reading to Top or Bottom of Jaw Readings DC...

Page 6: ...hown in Figure 2 600V i1010 AC DC CURRENT CLAMP OUTPUT 1mV A CAT III CAT III A ZERO 600V 600A 1000A Power On Indicator Power On Off Button ZERO Knob VR4 VR5 VR1 VR6 VR3 Decal el004f eps Figure 2 Calib...

Page 7: ...n 0 2 199 6 to 200 4 mV 7 Remove the Current Clamp jaws from the current coil and check that the DMM reads 000 0 If nec essary adjust the ZERO knob to obtain 000 0 8 Apply 4A dc 200 ampere turns to th...

Page 8: ...operation Replacement Parts Replacement parts are listed in Table 5 and shown in Figure 3 To obtain replacement parts call Fluke Corporation at 1 800 526 4731 Table 5 Replacement Parts Part Descriptio...

Page 9: ...R O A Case Bottom Jaw Spring Power ON OFF Button Poron Ring Brass Cable Clamp Cable Output Assembly Zero Knob Battery Door Top Case Decal Jaw Assembly AC D C CURRENT CLAM P O UTPUT 1m V A i1010 6 0 0...

Reviews: