6

Calibration and Adjustment

Always complete the Performance Test prior to Calibration. If Performance Test results are satisfactory,
calibration is not necessary. Otherwise, to ensure that the Current Clamp performs to specification, cali-
brate the Battery Indicator, DC Zero, and DC Amps and check the AC Amps function on a yearly basis.

When it is needed, be sure to complete the full calibration procedure (Battery Indicator, DC Zero, DC
Amps Calibration and AC Amps Check.) Partial completion of this procedure will not ensure that the
Current Clamp performs to specification.

Adjustment points are shown in Figure 2.

600V

i1010

AC/DC CURRENT CLAMP

OUTPUT 1mV/A

CAT III

CAT 

III

A

ZERO

600V
600A

1000A

~

Power On Indicator

Power On/Off Button

ZERO Knob

VR4
VR5
VR1
VR6

VR3

Decal

el004f.eps

Figure 2. Calibration Adjustment Points

Calibrating the Battery Indicator

Use the following procedure to calibrate the low battery indicator:

1.

 

Set the DC voltage generator to 9.0V dc.

2.

 

On the Current Clamp, disconnect the battery from the battery clip.

3.

 

Connect the DC voltage generator to the battery snap.

4.

 

Connect the Current Clamp output cable to a DMM. Check that the DMM reads 9.0V dc.

5.

 

On the Current Clamp, press the green power button on and check that the LED is on.

6.

 

Set the DC voltage generator to 7.0V dc. (The LED may be on or off at this point.)

7.

 

Adjust VR4 on the Current Clamp to the point where the LED cycles from on to off.

Summary of Contents for FLUKE i1010

Page 1: ...p use only specified replacement parts Caution If the Current Clamp is to be Calibrated you must have a replacement decal See the Replacement Parts section for further details This Current Clamp conta...

Page 2: ...Product 240 000 Storage Temperature 20 to 60 C 4 to 140 F Output Signal 1 mV per amp dc or ac Relative Humidity 0 to 95 0 to 30 C 0 to 75 30 to 40 C 0 to 45 40 to 50 C Working Voltage 600V CAT III ma...

Page 3: ...n Battery LED Test Complete the following procedure to verify that the Battery LED threshold is set correctly 1 Turn on the Calibrator and allow it to warm up and stabilize for 30 minutes 2 Place the...

Page 4: ...ollowing 1 Calibration necessitates removal of the decal on the case top Plan to install a new decal when cali bration is complete See Replacement Parts for decal ordering information 2 Turn on the Ca...

Page 5: ...104 0 97 5 102 5 12 2 0A 1 kHz 100 0A ac 100 0 mV NA NA 96 5 103 5 DC Amps Balance Test 13 4 0A dc 200 0A dc 200 0 mV 1 0A maximum change from Center of Jaw Reading to Top or Bottom of Jaw Readings DC...

Page 6: ...hown in Figure 2 600V i1010 AC DC CURRENT CLAMP OUTPUT 1mV A CAT III CAT III A ZERO 600V 600A 1000A Power On Indicator Power On Off Button ZERO Knob VR4 VR5 VR1 VR6 VR3 Decal el004f eps Figure 2 Calib...

Page 7: ...n 0 2 199 6 to 200 4 mV 7 Remove the Current Clamp jaws from the current coil and check that the DMM reads 000 0 If nec essary adjust the ZERO knob to obtain 000 0 8 Apply 4A dc 200 ampere turns to th...

Page 8: ...operation Replacement Parts Replacement parts are listed in Table 5 and shown in Figure 3 To obtain replacement parts call Fluke Corporation at 1 800 526 4731 Table 5 Replacement Parts Part Descriptio...

Page 9: ...R O A Case Bottom Jaw Spring Power ON OFF Button Poron Ring Brass Cable Clamp Cable Output Assembly Zero Knob Battery Door Top Case Decal Jaw Assembly AC D C CURRENT CLAM P O UTPUT 1m V A i1010 6 0 0...

Reviews: