![TensCare X-PR13 Instructions For Use Manual Download Page 3](http://html1.mh-extra.com/html/tenscare/x-pr13/x-pr13_instructions-for-use-manual_1087661003.webp)
Caution:
If tissue irritation occurs, discontinue treatment immediately.
Ask
your healthcare professional for advice before continuing further treatment
to prevent injury.
Caution:
Do not use a silicone-based lubricant on the metal plates of the
probe as it may decrease the effectiveness of the electrical stimulation.
Warning:
Do not use if the pack is opened or damaged.
INSTRUCTIONS FOR USE
1. Clean the probe with an alcohol-free
anti-bacterial wipe or with warm soapy water.
Rinse under running water and dry thoroughly.
2. Connect the lead from the base of the unit to the lead in the probe. Push
the pin ends firmly into the pigtail ends of the probe lead.
3. Before using the probe, you will need to visit the toilet.
4. Lubricate the metal electrode surfaces and probe tip with a water-based
lubricant, such as TensCare Go Gel or water.
Caution:
Do not use a silicone-based lubricant on the
stimulation contacts as it may decrease the effectiveness of the
muscle stimulation.
5. Choose a comfortable position, such as lying down on your bed on
your side with your knees raised.
Warning:
Ensure the unit is switched OFF before inserting the
probe.
6.
Anal insertion:
After wires are securely connected, insert the probe
into the anus whilst ‘bearing down’ (as in the action of passing stool)
to a comfortable limit until the base of the flange on the probe touches
the anus. The metal parts conduct the electrical pulse and should be
in contact with the main part of the muscle at all times. The tissues
close to the entrance are more sensitive, so you should avoid
stimulating them. It is recommended that the probe is inserted past
the sphincter muscles of the anus, unless directed otherwise by a
healthcare professional.
For Faecal incontinence, the aim is to stimulate the external sphincter
and/or pubo-rectal muscle, so circular electrodes should be placed
so that the external ring is just inside the sphincter.
For Urinary Stress incontinence, the aim is to stimulate the levator
muscles and the probe should be inserted deeper.