17
Square electrode pads (
E-CM5050)
should be used for all programmes
excluding
FACE 1
and
FACE 2.
Note:
These programmes must be used in combination with a healthy
diet and regular active exercises.
8.2.
PROGRAMMES SETTINGS
PROGRAMME 1: Facial Quick Tone
Phase
1
2
3
4
5
6
7
Total
Time (min)
20
20
Stimulation (s)
0
Pause (s)
1
Frequency (Hz)
70
Pulse width (µs)
100
Ramp (s)
1/1
PROGRAMME 2: Facial Deep Tone - Microcurrent
Phase
1
2
3
4
5
6
7
Total
Time (min)
2
16
8
26
Waveform
- monopolar
Sine Rect Rect
Frequency (Hz)
20
8
0.8
Pulse width (ms)
25
60
120
Intensity (µA) Fixed
300
250
80
PROGRAMME 3: Abdominal
Phase
1
2
3
4
5
6
7
Total
Time (min)
5
5
10
8
7
8
2
45
Stimulation (s)
2
4
8
10
6
4
1.2
Pause (s)
2.5
4
6
1
4
4
.5
Frequency (Hz)
90
50
10
4
30
65
100
Pulse width (µs)
200
300
500
600
360
300
80
Ramp (s)
.5
.5
.9
1.3
.6
.5
.2
Summary of Contents for Uniglo